products

Bante904 Benchtop Conductivity/Dissolved Oxygen Meter

short description:

High-accuracy multiparameter water quality meter contains 13 options, the meter is suitable for measuring the conductivity, TDS, salinity, resistivity and dissolved oxygen.


Product Detail

Product Tags

Conductivity/TDS/Salinity/Resistivity Modes

• Multiparameter water quality meter is equipped with a large backlit LCD display.

• 1 to 5 points calibration with automatic recognition for conductivity standards.

• Selectable cell constant (K=0.1/1/10), reference temperature (20/25°C), TDS factor, linear and pure water compensations, seawater and practical salinity measurement modes.

15412180546256

Other Features

• Automatic temperature compensation ensures accurate readings over the entire range.

• Auto-Read function senses and locks the measurement endpoint.

• Calibration due alarm prompts user to calibrate the meter regularly.

• Setup menu allows setting the number of calibration points, stability criteria, temperature unit, date and time, auto-power off, etc.

• Reset function automatically resumes all settings back to the factory defaults.

• Expanded memory stores or recalls up to 500 data sets.

• USB communication interface for data transfer and timed interval readings.

Specifications

   Model Bante904
   Conductivity  
   Range 0.01 to 20.00, 200.0, 2000µS/cm, 20.00, 200.0mS/cm
   Resolution 0.001, 0.01, 0.1, 1
   Accuracy ±0.5% F.S.
   Calibration Points 1 to 5 points
   Calibration Solutions 10µS, 84µS, 1413µS, 12.88mS, 111.8mS
   Temperature Coefficient Linear (0.0 to 10.0%/°C), pure water
   Reference Temperature 20/25°C
   Cell Constant K=0.1, 1, 10 or custom
   TDS  
   Range 0.00 to 10.00, 100.0, 1000ppm, 10.00, 200.0ppt
   Resolution 0.01, 0.1, 1
   Accuracy ±1% F.S.
   TDS Factor 0.1 to 1.0 (default 0.5)
   Salinity  
   Range 0.00 to 80.00ppt, 0.00 to 42.00psu
   Resolution 0.01
   Accuracy ±1% F.S.
   Resistivity  
   Range 0.00 to 20.00MΩ
   Resolution 0.01
   Accuracy ±1% F.S.
   Dissolved Oxygen  
   DO Range 0.00 to 20.00mg/L
   Resolution 0.01, 0.1mg/L, selectable
   Accuracy ±0.2mg/L
   % Saturation of Oxygen 0.0 to 200.0%
   Resolution 0.1, 1%, selectable
   Accuracy ±2.0%
   Calibration Points 1 or 2 points
   Pressure Correction 60.0 to 112.5kPa/450 to 850mmHg, manual
   Salinity Correction 0.0 to 50.0g/L, manual
   Temperature  
   Range 0 to 105°C/32 to 221°F
   Resolution 0.1°C/0.1°F
   Accuracy ±0.5°C/±0.9°F
   Offset Calibration 1 point
   Calibration Range Reading ±10°C
   General Specifications  
   Temperature Compensation 0 to 100°C/32 to 212°F, manual or automatic
   Stability Criteria Low or high
   Calibration Due Alarm 1 to 31 days or off
   Memory Stores up to 500 data sets
   Output USB communication interface
   Connector 6-pin nimi-DIN, 3.5mm jack socket
   Display Custom LCD (125×100mm)
   Power DC5V, using AC adapter, 220V/50Hz
   Dimensions 210(L)×188(W)×60(H)mm
   Weight 1.5kg

 


  • Previous:
  • Next:

  • Write your message here and send it to us